Repeat
Independent result under declared repeatability conditions.
Unit Converters
Analyze same-condition density repeats without confusing repeatability with accuracy. Every record is retained after unit normalization and the statistics are calculated from unrounded values.
Density repeatability study
Convert same-condition repeated density measurements to one unit and calculate mean, sample standard deviation, RSD, range, and Type A uncertainty of the mean.
| Repeat | Source value | Converted density | Deviation from mean |
|---|
How to use
Convert same-condition repeated density measurements to one unit and calculate mean, sample standard deviation, RSD, range, and Type A uncertainty of the mean.
Independent result under declared repeatability conditions.
Center of accepted repeated values.
Scatter estimate using n−1 degrees of freedom.
Standard deviation divided by the absolute mean.
Standard uncertainty of the mean estimated as s/√n.
Result interpretation
A small RSD indicates close repeats relative to the mean, not agreement with a certified value. Type A uncertainty becomes smaller as repeat count grows under stable conditions, but systematic effects, temperature bias, calibration, and sample inhomogeneity remain outside this statistic.
Calculation method
Parse every numeric record, reject non-positive or malformed entries, convert through kg/m³, calculate the unrounded mean, use the n−1 sample variance, divide s by |mean| for RSD, divide s by √n for Type A uncertainty, and preserve each row deviation.
Evidence checks
Operator, instrument, method, time interval, temperature, and sample handling should meet the intended repeatability definition.
A drifting bath can create real density change that appears as instrument imprecision.
Settling solids, bubbles, or evaporation can make sequential readings represent different measurands.
Do not delete an inconvenient value automatically; apply the documented method and retain exclusions with reasons.
Repeated identical digits may reflect display resolution rather than negligible underlying scatter.
Comparison with a certified reference and bias correction require separate evidence.
Type A is one component; include calibration, temperature, buoyancy, repeatability design, and reference values where applicable.
Visual explanation
Each repeat remains an individual marker. A center line shows the mean and a band shows one sample standard deviation, so clustering and time order remain visible instead of collapsing into one card.
Detailed calculation process
ρ̄ = Σρi/n; s = sqrt(Σ(ρi − ρ̄)²/(n − 1)); RSD = 100s/|ρ̄|; uA(ρ̄) = s/√n
| Symbol | Meaning | Required unit |
|---|---|---|
| ρi | converted repeat | target density unit |
| n | accepted repeat count | count |
| ρ̄ | arithmetic mean | target density unit |
| s | sample standard deviation | target density unit |
| RSD | relative scatter | % |
| uA | Type A standard uncertainty of mean | target density unit |
Reconciliation:Waiting for current inputs.
Defaults and assumptions
The default series is a compact demonstration near 850 kg/m³. It is not a repeatability limit or a certified data set.
| Check | Current value A | Current value B | Decision role |
|---|
Decision analysis
Use the output to describe observed repeat scatter and plan follow-up. Do not declare a method precise enough until the governing repeatability requirement, sample design, and uncertainty budget are compared.
Examine the ordered residual pattern before accepting a single standard deviation as the complete precision story. Alternating values can indicate digital resolution, a monotonic sequence can indicate thermal drift or settling, and clusters can reveal separate preparation states. Plot or review the records in acquisition order and compare the observed behavior with the method’s repeatability conditions. If an outlier rule is authorized, apply it to the original measurements, document the statistic and critical value, and retain both included and excluded results; never delete a row merely to improve RSD. Consider whether repeated readings are genuinely independent: rereading the same filled cell or stable instrument display may understate preparation and sampling variability. For a reported mean, combine Type A uncertainty with calibration, reference-material, temperature, resolution, buoyancy, and other justified components under the laboratory method. Compare the resulting expanded uncertainty with the decision requirement separately. A low calculated Type A value cannot compensate for a biased instrument, changing sample, or unsuitable method. Precision acceptance should therefore name the governing limit, degrees of freedom, repeat design, and whether individual results or the mean is the intended measurand.
Evidence and data lineage
Retain every raw reading, sequence, unit, temperature, instrument, operator, sample preparation, accepted exclusions, calculation version, and unrounded statistics.
Limits and exclusions
The page does not test normality, detect drift, calculate reproducibility, correct bias, apply an outlier test, or build a complete measurement uncertainty budget.
Reliable sources
Worked cases
Six stable-temperature repeats show low scatter, then a certified reference is used separately to assess bias.
A downward trend across repeats suggests settling; the team changes mixing and sampling rather than reporting one pooled SD.
Important note
Precision cannot establish accuracy. A tightly clustered set may still be wrong by a common systematic amount.
Sample standard deviation is undefined with only one observation.
It estimates population scatter from a sample whose mean was also estimated.
No. Convert or separate the records before entry.
No. Accuracy requires reference evidence and bias assessment.
Order can reveal drift, settling, warming, or carryover.
No. Separators without numbers do not create artificial records.
Only under the governing documented rule, with the exclusion retained.
Only the statistical contribution estimated from these repeats.
No. Repetition reduces mean scatter but not common bias.
Not with repeatability alone; use a reproducibility design.
RSD is dimensionless and displayed as percent.
Range depends strongly on extremes and repeat count.
The page uses unrounded converted values internally.
When the sample state changes or the sequence is not one stable population.