Lower range value
Density assigned to the lower signal endpoint.
Unit Converters
Translate a current, voltage, frequency, or other linear sensor signal into engineering density while preserving both signal and density endpoints and clearly flagging extrapolation.
Linear transmitter calibration
Use two documented signal and density endpoints to calculate a linear transfer function, live density, span fraction, sensitivity, and out-of-range status.
| Calibration point | Signal | Density | Interpretation |
|---|
How to use
Use two documented signal and density endpoints to calculate a linear transfer function, live density, span fraction, sensitivity, and out-of-range status.
Density assigned to the lower signal endpoint.
Density assigned to the upper signal endpoint.
Live signal position between signal endpoints.
Density change per signal unit.
Linear result outside the calibrated signal span.
Result interpretation
A fraction from 0 to 1 is inside the declared signal span. Values outside that interval are extrapolated even if the arithmetic remains finite. Sensitivity may be negative for an intentionally reversed density mapping; do not silently reorder endpoints.
Calculation method
Subtract lower signal from the live signal and divide by signal span. Multiply that fraction by density span and add the lower density endpoint. Calculate sensitivity as density span divided by signal span and preserve the original orientation.
Evidence checks
Endpoints must belong to the same transmitter range, configuration revision, and output mode.
Account for loop offsets, scaling in acquisition hardware, or digital raw counts before applying this mapping.
A linear transfer function is valid only when the sensor or transmitter output is configured linearly in density.
A falling density with rising signal is allowed mathematically but should be confirmed against configuration.
Out-of-range output may represent overrange, fault current, or saturation rather than a valid density.
A midpoint or certified-fluid check can reveal endpoint transcription or configuration error.
Correct signal mapping does not guarantee the sensor is free of coating, bubbles, temperature bias, or installation effects.
Visual explanation
Signal runs horizontally and mapped density vertically. Endpoint markers define the calibration line; the live marker remains on the line but changes status when it leaves the declared signal interval.
Detailed calculation process
fraction = (S − SLRV)/(SURV − SLRV); ρ = ρLRV + fraction(ρURV − ρLRV); sensitivity = (ρURV − ρLRV)/(SURV − SLRV)
| Symbol | Meaning | Required unit |
|---|---|---|
| S | live corrected signal | selected signal unit |
| SLRV,SURV | signal endpoints | selected signal unit |
| ρLRV,ρURV | density endpoints | selected density unit |
| fraction | signal span position | dimensionless |
| sensitivity | density per signal | density/signal unit |
| ρ | mapped density | selected density unit |
Reconciliation:Waiting for current inputs.
Defaults and assumptions
Defaults demonstrate a 4–20 mA loop mapped from 600 to 1200 kg/m³ with a 12 mA live signal. Confirm the actual transmitter configuration before use.
| Check | Current value A | Current value B | Decision role |
|---|
Decision analysis
Use an in-range mapped value for monitoring after calibration and installation checks. Route extrapolated or fault-region signals for instrumentation review instead of treating them as valid process density.
Validate the complete signal chain rather than only the endpoint arithmetic. A control-system tag can apply its own raw-count scaling, square-root extraction, offset, damping, or engineering conversion before the value reaches this page. Confirm where the entered live signal is measured and ensure the two calibration endpoints refer to that same stage. Compare the mapped midpoint with an independent configured checkpoint, because swapped density endpoints or an obsolete transmitter range can still produce a smooth line. Interpret overrange according to device documentation: some transmitters continue measuring beyond nominal range, while others use specific high or low currents for fault status. Do not report a numerically extrapolated density when the signal actually represents fault, saturation, open circuit, or diagnostic mode. If sensitivity is negative, confirm that the reverse orientation is intentional and preserved through alarms and displays. Correct transfer-function mapping also does not remove process error from coating, bubbles, vibration, temperature, pressure, installation, or sample nonrepresentativeness. Retain transmitter tag, range revision, loop location, signal quality, and checkpoint evidence with every operational use of the mapped density.
Evidence and data lineage
Retain transmitter tag, serial number, configuration revision, signal type, endpoint units, density range, live timestamp, correction chain, check-point evidence, and extrapolation status.
Limits and exclusions
The model excludes nonlinear sensor response, temperature compensation, digital protocol status, loop fault conventions, damping, calibration uncertainty, and process installation effects.
Reliable sources
Worked cases
The 12 mA signal maps to the midpoint density and agrees with a certified-fluid checkpoint.
A 22 mA input extrapolates numerically but is routed to instrument diagnostics instead of reported as density.
Important note
A valid equation can translate an invalid instrument signal. Interpret device status and configured fault behavior before using the mapped number.
Yes, but confirm the reversed configuration intentionally.
The transfer function is undefined and the inputs are rejected.
No. Signal labels are generic; all three signals must share one unit.
Yes for a documented linear voltage output.
Yes when endpoint counts and preprocessing are documented.
The signal is halfway between declared endpoints.
No. It is extrapolated and may be overrange or fault.
It preserves an intentionally reversed endpoint orientation.
No. It applies entered calibration endpoints.
No. Use the transmitter’s validated compensation or a separate model.
Two correct-looking endpoints can still be transcribed from the wrong range.
No. Unrounded fraction controls extrapolation.
Only if both endpoints and output are explicitly specific gravity.
It records endpoints, live signal, mapped density, sensitivity, and range status.